Sie befinden sich hier: Startseite » Produkte » GJS-P1060 Jacobsen Germinator

GJS-P1060 Jacobsen Germinator

Product Information

  • Temperature 0 °C to +50°C, optionally -30 °C to +80 °C
  • Humidity Range up to 99% r. h.
  • Test Room 1060 Litres
  • 10 Levels for Germination acc. to the Jacobsen Method TOP
  • Program Controller Control2015touch
  • Screen Recorder and USB/Ethernet
  • Program-controlled, dimmable LED Lighting
  • Optional Humidity
TypeGJS-P1060
Test Room
Volume1060 l
Temperature Range0 bis +50 °C
Humidity Rangebis 99% rel. F.
Deviation in Time±0,5 °C
Height1500 mm
Width2 x 610 mm
Depth585 mm
Number of shelves 10
Maximum load per shelf25 kg
Appliance
Height2105 mm
Width1460 mm
Depth980 mm
Electr. Connection230 V / 50 Hz
Net weight of the standard unit300 kg
Refrigerant
Type / GWPR290 / 3
CO2-Equivalent< 0,001 t

Enquiry Questionnaire

Please choose the options, which are relevant for the product enquiry.

Temperature Options

test

Extension of the minimum temperature to -20 °C.

Humidity Options

Ion exchanger for direct connection to the drinking water pipe, if demineralized water supply is not available.

Canister for water supply at installation places without water connection.

Optional Accessories

A 45 mm duct in the lateral panel allows insertion of measuring lines and hoses into the test room. Delivery including cover.

Gas-tight closing plug, matching to the tubular port. The gasket system with a flexibility in the range of 4 mm to 23 mm is based on a rubber module with removable skins and can be adapted to the cables, tubes etc. in different sizes.

Test room socket for connection of shakers, magnetic mixers etc. in the test room. Switching-on times can be freely programmed by means of the CONTROL2015 touch.

Export execution for power supply 110 V/60 Hz.

Communication and Software

Additional measuring sensor PT1000 for temperature control and recording of specimen.

Analogue output test room temperature for external recorder Signal 0-10 V, 0-20 mA, 4-20 mA

Analogue output test room humidity for external recorder Signal 0-10 V, 0-20 mA, 4-20 mA

- Monitoring of up to 32 appliances   - Compatible with Control 2000 and Control2015 touch   - Data logging and graph representation   - Logging of door openings with time stamp   - Logging of alarms with time stamp   - Automated daily print is possible   - Weekly and monthly print   - Archiving per day and control circuit   - Automated backup on server drive  

- Monitoring of appliances with Control2015 touch - Data safety in conformity with the guidelines 21CFR Part 11 - All events are recorded in a digital logbook, inseparably connected with the recorded data. - Signature in connection with the user management of the appliances - Automated prints  - Automated backup on server drive

Calibration and Qualification

Calibration on site with 9 spatially distributed sensors with a temperature value.

Calibration on site with 9 spatially distributed sensors with a temperature/humidity value.

Calibration on site with 9 spatially distributed sensors with a temperature value.

Calibration on site with 9 spatially distributed sensors with a temperature/humidity value.

We are always ready to support the validation of your appliances. If desired, the validation can be executed by our company. The validation includes the necessary measurements, as well as any documents, such as DQ, IQ, OQ and PQ, manuals, wiring diagrams, plans, appliance logbook, manufacturer certificates for the sensors and for the most important components. If required, we are also at your disposal for the annual recalibration.

Execution and documentation of the design qualification, qualification of installation and operation qualification including calibration on site with 9 spatially distributed sensors with a temperature value on site.

Execution and documentation of the design qualification, qualification of installation and operation qualification including calibration on site with 9 spatially distributed sensors with a temperature/humidity value on site.

Execution and documentation of the performance qualification including calibration on site with 9 spatially distributed sensors with a temperature value on site.

Execution and documentation of the performance qualification including calibration on site with 9 spatially distributed sensors with a temperature/humidity value on site.

    Title*

    Academic Title

    First Name*

    Surname*

    Company / Institute*

    Department

    Street*

    House Number*

    ZIP*

    City*

    Country*

    E-Mail-Adress*

    Phone*

    Newsletter

    Yes, I want to receive RUMED's free newsletter. I can unsubscribe easily at any time.

    Request catalogue

    Subject

    Info

    Please enter the unit type (1) and the serial number (2) into the enquiry form.

    BUnit type and serial number of the Control 2000 are displayed after actuation of the key i.

    Your Message